LORA T3: Data logger with display and wireless data transmission for ultra-low freezers, LN2 tanks and stoves

Lora T3 is a data logger with display, designed for ultra-low freezers, LN2 tanks and stoves, equipped with an external PT100 probe with Teflon cable. The logger has a measuring range from -200°C to +150°C.

It stores up to 10,000 measurements and automatically transmits data to the LoRa Gateway, via 868 MHz radio communication, with range up to 10Km in open field. The data is then automatically transmitted to the MySirius software in the cloud or locally.

Data sheet – EN

Measured size: Temperature
Sensor: 100 Ω Platinum RTD Class A (external) Ø 3×20 mm with 6m Teflon cable
Measuring range: -200 ÷ +150°C
Resolution: 0,01°C
Accuracy: ± 0,2°C from -20°C to 0°C  |  ± 0,5°C over
Storage: 10.000 readings
Output: Radio communication 868 MHz
Software: MySirius (cloud or local)
IP Rating: IP65
Display: Display and LED for alarm status recognition
Battery: 3.6 V lithium battery, replaceable
Fixing: Integrated magnetic fixing + support for wall fixing
Certification: It is possible to request a LAT certificate of calibration